Table 1 Recombinant HRP2 concentrations used in Trial 1, and equivalent parasite densities derived from the median HRP2 concentration of dilutions of 200 parasite/μL of the WHO global specimen bank

From: Laboratory demonstration of a prozone-like effect in HRP2-detecting malaria rapid diagnostic tests: implications for clinical management

HRP2 concentration (ng/mL)

Line intensity (densitometer)

Mean line intensity (3 readers)

Calculated parasite density equivalent (parasite/μL)

15

1.8

1.3

313

150

7.2

3.0

3,125

1,500

16.9

4.0

31,250

15,000

17.9

4.0

312,500

150,000

13.0

3.7

3,125,000

300,000

10.8

3.0

6,250,000

500,000

5.2

2.3

10,416,667

750,000

4.4

2.0

15,625,000

1,000,000

2.4

2.0

20,833,333

1,500,000

1.5

1.0

31,250,000