Items addressed in procedure section | Number (%) |
---|---|
Bring the RDT device and buffer to room temperature | 32 (80.0) |
Check the integrity of the device package | 9 (22.5) |
Check expiry date | 27 (67.5) |
Use the device immediately after opening | 28 (70.0) |
Place the device on a level surface | 0 (0.0) |
Check the desiccant for signs of exposure to humidity | 11 (27.5) |
Write down sample identification | 3 (7.5) |
Wipe finger with alcohol | 26 (65.0) |
Allow the finger to dry before pricking | 12 (30.0) |
Hold the transfer device (loop, straw) vertical | 8 (20.0) |
Hold the buffer vial vertical | 12 (30.0) |
Do not to use another buffer than the one provided with the kit | 9 (22.5) |
Use an adequate light source for reading | 3 (7.5) |
Items addressed in interpretation section | Number (%) |
All possible line combinations for invalid test results are mentioned | 12 (30.0) |
All possible test line combinations for positive test results are mentioned | 31 (77.5) |
Interpretation of a faint test line as positive is mentioned | 8 (20.0) |
Causes of false negative results are mentioned, in particular low parasite densities | 11 (27.5) |
Causes of false positive results are mentioned, e.g. presence of the rheumatoid factor | 3 (7.5) |
Persistence of HRP-2 is mentioned | 19 (47.5) |
To repeat the test in case of a negative RDT result and persistent suspicion of malaria is mentioned | 1 (2.5) |