Table 5 Parasite counting results for the 13 positives slides of the EGMIS

From: Performance of a malaria microscopy image analysis slide reading device

   CMD WHO std Confidence Interval  
EGMIS (+) Machine Counts MEDIAN 25% below 25% above 99% lower limit 99% upper limit Correct
44 100 1,151 863 1,439 247 2,050  
56 800 675 506 844 429 1,049
61 2,800 673 505 841 419 1,006  
63 100 47 35 59 22 96  
68 8,000 1,916 1,437 2,395 1,056 3,176  
72 1,600 652 489 815 156 1,178  
74 150 110 83 138 47 168
78 50 207 155 258 24 301
115 31,600 34,811 26,108 43,514 24,025 47,726
117 17,200 54,613 40,960 68,266 38,770 75,645  
118 571 724 543 905 392 1,225
137 1,500 596 447 745 305 1,328  
143 10,000 160 120 200 125 226