Skip to main content

Advertisement

Table 9 Critical assay parameters

From: Validation of an enzyme-linked immunosorbent assay for the quantification of human IgG directed against the repeat region of the circumsporozoite protein of the parasite Plasmodium falciparum

LOD 0.2 EU/ml
LOQ 0.3 EU/ml
Cut-off 0.5 EU/ml
Precision (CV) 22.46%
Linearity 1.1 - 2440 EU/ml
Analytical range 0.3 - 2440 EU/ml
  1. LOD, limit of detection.
  2. LOQ, limit of quantification.
  3. CV, coefficient of variation.