Skip to main content

Table 2 Deki Reader and Visual diagnostic performance using microscopy asgold standard

From: Clinical performance of an automated reader in interpreting malaria rapid diagnostic tests in Tanzania

  N Sensitivity (95% CI) Specificity (95% CI) PPV (95% CI) NPV (95% CI) Overall diagnostic accuracy (95% CI
Deki reader 1346 95.3 (90.6, 97.7) 94.9 (93.5, 96.1) 70.3 (63.7, 76.2) 99.4 (98.7, 99.7) 95 (93.7, 96.0)
Visual 1346 94.7 (89.8, 97.3) 95.6 (94.2, 96.6) 72.8 (66.2, 78.6) 99.3 (98.6, 99.6) 95.5 (94.2, 96.5)
Comparison   0.7 (-2.7, 4.2) -0.7 (-1.5, 0.1)