Skip to main content

Table 2 Limit of detection for pLDH in three rapid diagnostic tests by using artificially diluted samples of Plasmodium vivax

From: pLDH level of clinically isolated Plasmodium vivax and detection limit of pLDH based malaria rapid diagnostic test

pLDH (ng/mL)

Parasitaemia(/μL)

SD Bioline (n = 5)

OptiMAL (n = 5)

Humasis(n = 5)

5,000

8,700.0

100%

100%

100%

500

870.0

100%

100%

100%

250

435.0

100%

100%

100%

50

87.0

100%

100%

100%

25

44.0

100%

0%

100%

5

8.7

0%

0%

0%

0.5

0.9

0%

0%

0%

0

0

0%

0%

0%