From: Field evaluation of diagnostic performance of malaria rapid diagnostic tests in western Kenya
 | Sensitivity %, 95 % CI | Specificity %, 95 % CI | Predictive value | Likelihood ratio | Methods agreement % | κ, 95 % CI | ||
---|---|---|---|---|---|---|---|---|
Positive test %, 95Â %Â CI | Negative test %, 95Â %Â CI | Positive test %, 95Â %Â CI | Negative test %, 95Â %Â CI | |||||
BinaxNOW 665-025 | 71.9 | 84.8 | 83.3 | 74.2 | 4.7 | 0.3 | 78.2 | 0.6 |
65.9–77.3 | 79.7–89.1 | 77.7–87.9 | 68.6–79.2 | 3.5–6.4 | 0.3–0.4 | 0.5–0.6 | ||
SD Bioline 05FK60 | 75.4 | 81.1 | 80.4 | 76.1 | 3.9 | 0.3 | 78.0 | 0.6 |
69.6–80.5 | 75.6–85.7 | 74.8–85.2 | 70.5–81.2 | 2.9–5.1 | 0.3–0.4 | 0.5–0.6 | ||
SD Bioline 05FK50 | 74.8 | 80.8 | 80.3 | 75.4 | 4.0 | 0.3 | 79.0 | 0.6 |
68.9–80.5 | 75.2–85.6 | 74.6–85.2 | 69.6–80.54 | 3.1–5.3 | 0.3–0.4 | 0.5–0.7 | ||
First Response I16FRC | 71.1 | 84.4 | 82.7 | 73.6 | 4.6 | 0.3 | 77.6 | 0.6 |
 | 65.1–76.6 | 79.3–88.7 | 77.1–87.5 | 68.0–78.6 | 3.4–6.2 | 0.3–0.4 | 0.5–0.6 | |
First Response I13FRC | 73.8 | 82.8 | 81.8 | 75.1 | 4.3 | 0.3 | 78.2 | 0.6 |
67.9–79.1 | 77.5–87.3 | 76.2–86.6 | 69.4–80.1 | 3.2–5.7 | 0.3–0.4 | 0.5–0.6 | ||
CareStart G0131 | 73.8 | 82.4 | 81.5 | 75.0 | 4.2 | 0.3 | 78.0 | 0.6 |
67.9–79.1 | 77.0–86.9 | 75.7–86.3 | 69.4–80.1 | 3.2–5.5 | 0.3–0.4 | 0.5–0.6 | ||
CareStart G0181 | 72.3 | 81.6 | 80.4 | 73.7 | 3.9 | 0.3 | 76.8 | 0.6 |
66.4–77.7 | 76.1–86.2 | 74.7–85.4 | 68.0–78.9 | 2.9–5.2 | 0.3–0.4 | 0.5–0.6 | ||
CareStart G0141 | 75.4 | 82.4 | 81.8 | 76.1 | 4.3 | 0.3 | 78.8 | 0.5 |
69.6–80.5 | 77.0–86.9 | 76.3–86.5 | 70.5–81.2 | 3.2–5.7 | 0.3–0.4 | 0.5–0.6 |