TableĀ 4 Individual and institutional characteristics associated with accurate malaria microscopy diagnosis in surveyed health facilities in malaria low-transmission areas of Kenya, 2014

From: Factors associated with malaria microscopy diagnostic performance following a pilot quality-assurance programme in health facilities in malaria low-transmission areas of Kenya, 2014

Characteristic

Ā 

Slides n (%)

Accurate diagnosis n (%)

Unadjusted

Adjusted

Prevalence ratio

95% confidence interval

Prevalence ratio

95% confidence interval

Individual

Ā Recent microscopy refresher training

No

295 (39.0)

197 (66.8)

1.00 (Ref)

Ā 

1.00 (Ref)

Ā 

Yes

461 (61.0)

446 (96.7)

40.5

15.1ā€“108.6

13.8

4.6ā€“41.4

Ā More than diploma-level initial training

No

141 (18.7)

109 (77.3)

1.00 (Ref)

Ā Ā Ā 

Yes

615 (81.3)

534 (86.8)

2.1

0.9ā€“4.6

Ā Ā 

Ā ā‰„5Ā years of work experience

No

137 (18.1)

78 (56.9)

1.00 (Ref)

Ā 

1.00 (Ref)

Ā 

Yes

619 (81.9)

565 (91.3)

23.7

9.7ā€“57.6

3.8

1.5ā€“9.9

Ā Worked in malaria high-transmission area

No

383 (50.7)

290 (75.7)

1.00 (Ref)

Ā Ā Ā 

Yes

373 (49.3)

353 (94.6)

12.1

5.2ā€“28.3

Ā Ā 

Ā Knowledge of malaria diagnostic and treatment guidelines

No

194 (25.7)

119 (61.3)

1.00 (Ref)

Ā Ā Ā 

Yes

562 (74.3)

524 (93.2)

25.6

10.4ā€“62.9

Ā Ā 

Ā Knowledge of malaria epidemiology in county

No

94 (12.4)

45 (47.9)

1.00 (Ref)

Ā Ā Ā 

Yes

662 (87.6)

598 (90.3)

22.8

8.7ā€“60.3

Ā Ā 

Ā Knowledge of malaria cases importation

No

63 (8.3)

20 (31.7)

1.00 (Ref)

Ā Ā Ā 

Yes

693 (91.7)

623 (89.9)

21.1

7.2ā€“62.1

Ā Ā 

Institutional

Ā Quality-assurance pilot programme

No

378 (50.0)

292 (77.2)

1.00 (Ref)

Ā 

1.00 (Ref)

Ā 

Yes

378 (50.0)

351 (92.8)

6.0

1.9ā€“18.9

4.3

1.0ā€“11.0

Ā Good optical condition of microscope(s)

No

72 (9.5)

46 (63.9)

1.00 (Ref)

Ā Ā Ā 

Yes

684 (90.5)

597 (87.3)

7.6

1.1ā€“51.4

Ā Ā 

Ā Rural location

No

505 (66.8)

415 (82.2)

1.00 (Ref)

Ā Ā Ā 

Yes

251 (33.2)

228 (90.8)

2.7

0.7ā€“10.1

Ā Ā 

Ā Participation in SLIPTA program

No

647 (85.6)

539 (83.3)

1.00 (Ref)

Ā Ā Ā 

Yes

109 (14.4)

104 (95.4)

4.8

0.8ā€“28.8

Ā Ā 

Ā >3 laboratory staff

No

361 (47.8)

308 (85.3)

1.00 (Ref)

Ā Ā Ā 

Yes

395 (52.2)

335 (84.8)

1.1

0.3ā€“3.3

Ā Ā 

Ā Hospital-level facility

No

432 (57.1)

376 (87.0)

1.00 (Ref)

Ā Ā Ā 

Yes

324 (42.9)

267 (82.4)

0.8

0.2ā€“2.5

Ā Ā 

Ā WorkloadĀ >10 slides per day

No

269 (35.6)

239 (88.8)

1.00 (Ref)

Ā Ā Ā 

Yes

487 (64.4)

404 (83.0)

0.6

0.2ā€“2.1

Ā Ā 
  1. Italic denotes statistical significance; recent training was defined as in the year prior to the survey
  2. SLIPTA stepwise laboratory improvement towards accreditation, sponsored by World Health Organization; Ref reference