Skip to main content

Table 3 Performance of visual interpretation and Deki Reader™ of the SD Bioline Malaria Ag Pf/Pan RDT against microscopy

From: Evaluation of the Deki Reader™, an automated RDT reader and data management device, in a household survey setting in low malaria endemic southwestern Uganda

 

Microscopy

Sensitivity (95% CI)

p-value*

Specificity (95% CI)

p-value*

Positive

Negative

Visual interpretation

      

 Positive

15

23

88.2 (63.7, 98.5)

95.8 (93.8, 97.3)

 Negative

2

526

0.4

0.06

Deki Reader™

   

 Positive

16

24

94.1 (73.0, 99.0)

95.5 (93.4, 96.9)

 Negative

1

525

  
  1. *Reported p-value using McNemar’s test to assess difference in sensitivity between visual inspection and the Deki Reader™