Skip to main content

Table 2 Dilution linearity and accuracy of qSAT assay

From: Quantification of malaria antigens PfHRP2 and pLDH by quantitative suspension array technology in whole blood, dried blood spot and plasma

SamplePf pLDHPfHRP2
Expected concentration (pg/ml)% change% deviationExpected concentration (pg/ml)% change% deviation
115,62541.624.2781.332.823.4
278,12.518.317.4390.618.510.8
33906.30.917.1195.36.810.2
41953.1210.797.708.5
5976.65.610.448.85.210.1
6488.34.214.224.414.719
7244.15.514.912.212.727.9
8122.110.422.26.118.547
Overall11.116.413.619.6