Table 2 Calibration. Comparisons between the HPF calling performed by the model and the expert decision. \(^{*}\) HPF was detected only by the model but not confirm by the expert. \(^{**}\) HPF was not detected by the model but detected by the expert only. \(^{***}\) no FP nor FN on the whole sample

From: A timed tally counter for microscopic examination of thick blood smears in malaria studies

Calibration False positive rate\(^{*}\) False negative rate\(^{**}\) Full agreement\(^{***}\)
\(p=0.01\) \(\mu =10\) 20.18% (45/223) 1.35% (3/223) 9/20
\(p=0.05\) \(\mu =10\) 7.39% (13/176) 1.70% (3/176) 11/20
\(p=0.01\) \(\mu =15\) 11.67% (21/180) 2.78% (5/180) 9/20
\(p=0.05\) \(\mu =15\) 9.21 % (14/152) 6.58 % (10/152) 11/20