Table 3 Sensitivity and Specificity of microscopy and the different qPCR methods studied compared to LAMP Alethia® Malaria

From: Evaluation of two commercial kits and two laboratory-developed qPCR assays compared to LAMP for molecular diagnosis of malaria

 

Microscopy

Plasmodium Typage (Bio-Evolution)

Ampliquick® Malaria

In-house HRM

In-house TaqMan

 

 + 

 + 

 + 

 + 

 + 

LAMP + (N = 104)

86

18

99

5

101

3

102

2

104

0

LAMP − (N = 43)

0

43

0

43

0

43

0

43

0

43

Specificity

100%

100%

100%

100%

100%

Sensitivity 95% CI

85.2% [74–89.4]

95.4% [89.1–98.4]

97.2% [91.8–99.4]

98.1% [93–99.8]

100%[100–100]

Kappa 95% CI

0.74 [0.63–0.85]

0.92 [0.85–0.99]

0.95 [0.9–1]

0.97 [0.92–1]

1 [1–1]