Table 6 Comparison of the BD and Bender MedSystems Kits

From: Comparison of commercial kits to measure cytokine responses to Plasmodium falciparum by multiplex microsphere suspension array technology

Cytokine

Number of positive observations

Spearman's Rho

GM (pg/ml)

BD

GM (pg/ml)

Bender Medsystems

% difference in GM between 2 plates (95% CI)

Wilcoxon signed rank test

IFN-γ

23

0.75**

80.00

145.69

45 (62, 21)

0.043

IL-2

13

0.86**

80.87

113.10

29 (57, 19)

0.339

IL-6

70

0.94**

867.64

433.89

99.9 (72,133)

<0.0001

IL-10

20

0.88**

40.84

29.87

37 (-5, 96)

0.002

TNF

44

0.7**

36.83

56.64

35 (51, 13)

0.010

  1. Number of positive observation concordant between both assays, with the Spearman's Rho to test for significant correlations between the assays. **= Spearman's Rho test with significance of P < 0.01, * = Spearman's Rho test with significance of P < 0.05, NS = Spearman's Rho test which in non-significant. Geometric means of both assays with the % difference in geometric means. The table also shows Wilcoxon Rank Sum Significant differences in the quantity of cytokine detected in samples between the BD and Bender MedSystems kits. GM = geometric mean.