Table 4 Sensitivity by parasitaemia level of the SD 05FK60 rapid diagnostic test

From: Routine parallel diagnosis of malaria using microscopy and the malaria rapid diagnostic test SD 05FK60: the experience of Médecins Sans Frontières in Myanmar

Parasite density

P. falciparum

P. vivax/P. malariae

 

Number of RDT positives/Total positives by micoscopy

Sensitivity %

Number of RDT positives/Total positives by micoscopy

Sensitivity %

+

34/57

59.7

14/104

13.5

++

62/66

93.9

44/63

69.8

+++

134/148

90.5

153/159

96.2

++++

249/260

95.8

255/261

97.7

Total

479/531

90.2

466/587

79.4

Chi-square test for trend

p <0.05

p <0.05

  1. A total of 2,585 blood samples was examined at the Sittwe and Thetkalpyin clinics, Rakhine State, Myanmar.
  2. Estimated parasite densities: + = 10–90 p/μl, ++ = 100–1,000 p/μl, +++ = 1,000-10,000 p/μl, ++++ = >10,000 p/μl. N: total number of blood samples; n: total detected by SD Bioline Malaria Ag P.f/Pan (05FK60).