From: A bioassay method validation framework for laboratory and semi-field tests used to evaluate vector control tools

Stage

Notes

1. Select comparison method

 

2. Determine the maximum number of replicates that can be performed in a single day

Four replicates are the minimum to calculate within-run imprecision

3. Perform 30 replicates each of new method and comparator method in parallel over the smallest possible number of days

Note that for methods with extremely high variation, 30 replicates may be insufficient