- Research
- Open access
- Published:
Comparison of two malaria multiplex immunoassays that enable quantification of malaria antigens
Malaria Journal volume 21, Article number: 176 (2022)
Abstract
Background
Immunoassay platforms that simultaneously detect malaria antigens including histidine-rich protein 2 (HRP2)/HRP3 and Plasmodium lactate dehydrogenase (pLDH), are useful epidemiological tools for rapid diagnostic test evaluation. This study presents the comparative evaluation of two multiplex platforms in identifying Plasmodium falciparum with presence or absence of HRP2/HRP3 expression as being indicative of hrp2/hrp3 deletions and other Plasmodium species. Moreover, correlation between the malaria antigen measurements performed at these platforms is assessed after calibrating with either assay standards or international standards and the cross-reactivity among Plasmodium species is examined.
Methods
A 77-member panel of specimens composed of the World Health Organization (WHO) international Plasmodium antigen standards, cultured parasites for P. falciparum and Plasmodium knowlesi, and clinical specimens with mono-infections for P. falciparum, Plasmodium vivax, and Plasmodium malariae was generated as both whole blood and dried blood spot (DBS) specimens. Assays for HRP2, P. falciparum–specific pLDH (PfLDH), P. vivax–specific pLDH (PvLDH), and all human Plasmodium species Pan malaria pLDH (PanLDH) on the Human Malaria Array Q-Plex and the xMAP platforms were evaluated with these panels.
Results
The xMAP showed a higher percent positive agreement for identification of hrp2-deleted P. falciparum and Plasmodium species in whole blood and DBS than the Q-Plex. For whole blood samples, there was a highly positive correlation between the two platforms for PfLDH (Pearson r = 0.9926) and PvLDH (r = 0. 9792), moderate positive correlation for HRP2 (r = 0.7432), and poor correlation for PanLDH (r = 0.6139). In Pearson correlation analysis between the two platforms on the DBS, the same assays were r = 0.9828, r = 0.7679, r = 0.6432, and r = 0.8957, respectively. The xMAP HRP2 assay appeared to cross-react with HRP3, while the Q-Plex did not. The Q-Plex PfLDH assay cross-reacted with P. malariae, while the xMAP did not. For both platforms, P. knowlesi was detected on the PvLDH assay. The WHO international standards allowed normalization across both platforms on their HRP2, PfLDH, and PvLDH assays in whole blood and DBS.
Conclusions
Q-Plex and xMAP show good agreement for identification of P. falciparum mutants with hrp2/hrp3 deletions, and other Plasmodium species. Quantitative results from both platforms, normalized into international units for HRP2, PfLDH, and PvLDH, showed good agreement and should allow comparison and analysis of results generated by either platform.
Background
The standard of care for malaria diagnosis is blood smear microscopy and antigen detection through rapid diagnostic test (RDT). Microscopy has limitations in terms of difficulty in identifying mixed infections, and in user expertise and training requirements [1]. RDTs are more amenable for the diagnosis of malaria in settings with limited laboratory infrastructure where the majority of the malaria disease burden lies.
Several biomarkers, including Plasmodium falciparum–specific histidine-rich protein 2 (HRP2), Plasmodium lactate dehydrogenase (pLDH), and Plasmodium aldolase (pAldo), have been demonstrated to provide the discriminatory ability for detecting malaria parasites and classifying Plasmodium species [2]. In particular, the presence of a pan-epitope and species-specific epitopes on pLDH provides a tool to detect Plasmodium parasite as well as to classify the specific parasite species, such as P. falciparum and Plasmodium vivax, which are the major causes of human malaria. HRP2 has a high turnover rate during the asexual cycle and has an extended half-life in blood compared to pLDH, resulting in it being more abundant during malaria infection [3,4,5]. As such, the most widely used RDTs for P. falciparum target HRP2 and a homologous protein HRP3. HRP2 and HRP3 are nonessential proteins, and P. falciparum mutants with deletion of either or both genes coding for these proteins have been increasing in prevalence in malaria endemic countries, impacting the sensitivity and utility of HRP2-based tests in these settings [6].
The gold standard for malaria detection is confirmation of the presence of parasite DNA or RNA in whole blood by Polymerase chain reaction (PCR) testing. RDTs detect parasite antigens in whole blood, the presence of which does not fully correlate with that of the parasite nucleic acid. Quantitative enzyme-linked immunosorbent assays (ELISA), targeting the diagnostic malaria antigens, provide an independent approach for confirming the presence of malaria antigens in samples and therefore can inform the evaluation of RDTs [7, 8]. ELISA technologies that are capable of multiplexing offer many key qualities, including high-throughput potential, more results per sample, and lower sample volumes. Several laboratories have developed the tools to simultaneously quantify malaria antigens using two different technology platforms: the planar-based array and the magnetic bead-based platforms. Today there is only one commercial multiplexed assay for malaria antigen quantification, which is a planar array-based platform (the Q-Plex technology) and detects five biomarkers: HRP2, pan-specific pLDH (PanLDH), P. falciparum–specific pLDH (PfLDH), P. vivax–specific pLDH (PvLDH), and C-reactive protein (CRP) in whole blood and DBS [9, 10]. The bead-based platform using the xMAP technology has been applied for the development of two noncommercial multiplex malaria antigen assays targeting HRP2, PanLDH, PfLDH, PvLDH and pAldo in whole blood, plasma, and dried blood spot (DBS) [11,12,13].
In addition to identifying P. falciparum mutants with hrp2 deletion, these two multiplexed assays have been used for the investigation of the dynamics of antigen clearance, conducting malaria surveillance studies, anti-malarial drug clinical trials, and evaluating point-of-care tests [8, 14,15,16]. The different multiplexed assays use different antibody reagents as well as different calibration standards, resulting in a difference in overall performance and quantification across platforms. There is a recognition in the malaria community for the need to align assay results with international antigen standards. Recently, the National Institute for Biological Standards and Control (NIBSC) established lyophilized World Health Organization (WHO) international standards for P. falciparum and P. vivax antigens in order to ensure accurate results and the quality of malaria tests [17, 18]. This study sought to compare the performance of two multiplex assays for detecting malaria antigens in serially diluted samples (whole blood or DBS) that were prepared by spiking WHO international standards, parasite culture, or clinical samples with a view to verifying the compatibility of data between both platforms. The assays were the Human Malaria Array hosted on the Q-Plex platform (Quansys, Logan, UT, USA) and the multiplex bead-based assay on the xMAP platform developed by ISGlobal.
Methods
Reagents
The WHO international standards for P. falciparum (product code: 16/376) and P. vivax (product code: 19/116) antigens were purchased from the NIBSC (Hertfordshire, UK). Plasmodium falciparum W2 (hrp2+hrp3+) and Dd2 (hrp2–hrp3+) strains were obtained from Biodefense and Emerging Infections Research Resources Repository (BEI) Resources (Manassas, VA, USA) and the 3BD5 (hrp2–hrp3–) strain was obtained from the National Institute of Allergy and Infectious Diseases (Bethesda, Maryland, USA). Plasmodium knowlesi strain A1-H2 was a kind gift from Dr. Rob Moon (London School of Hygiene and Tropical Medicine, UK). Clinical specimens for P. falciparum, P. vivax, and Plasmodium malariae were acquired from Discovery Life Sciences (Santa Barbara, CA, USA). For calibration standards of the xMAP platform, HRP2 protein was purchased from Microcoat Biotechnologies’ (Starnberger See, Germany), and recombinant PfLDH and PvLDH proteins were purchased from MyBioSource (San Diego, CA, USA), Relia-Tech (Santa Fe Springs, CA, USA), and Microcoat. Pooled ethylenediaminetetraacetic acid–anticoagulated blood from five O + donors was used to prepare the sample panels.
Culture
Plasmodium falciparum and P. knowlesi laboratory strains were in vitro cultured according to procedures described previously [19, 20]. Synchronization of P. falciparum and P. knowlesi cultures was performed by D-sorbitol treatment or gradient centrifugation procedures using Nycodenz® solution (Axis-shield Diagnostics Ltd, Scotland) in 10 mM HEPES (pH 7.0), respectively [20, 21]. Parasitaemia was determined via staining of smear and light microscopy with a 100 × oil objective.
Sample panel preparation
The two-fold dilution series of 67 samples were prepared after spiking the material from the aforementioned antigen sources in pooled blood and aliquoting into cryovials, which were then stored at − 80 °C in a PATH laboratory until use. DBS samples were subsequently prepared from matched blood samples by spotting 60 µL of blood onto Whatman® 903 protein saver cards (GE Healthcare, Chicago, IL, USA), drying at room temperature overnight, and then storing at − 20 °C in sealed plastic pouches containing desiccant packets as described previously [10]. The sample panels consisting of 67 samples in frozen blood and DBS (Table 1), except those prepared with the WHO international standard P. vivax antigen, were sent to the ISGlobal laboratory, where the testing occurred. To limit the potential variability arising from differences in sample integrity that could happen during the transport, a set of the sample panels was subjected to a similar shipping time and storage conditions. Ten samples with WHO international standard P. vivax antigen were independently prepared in whole blood and DBS by each laboratory (Table 1).
Q-Plex platform procedure
The 5-Plex was run as per the manufacturer’s instructions. All whole blood and DBS samples were subjected to blinded experiments while performing the Q-Plex platform procedure. Whole blood samples were directly used without any manipulation, while the eluates were used for DBS by first incubating a 6 mm disc punched out from each DBS in elution buffer overnight as described previously [10]. Calibrators and samples were prepared according to the manufacturer’s protocol. Each sample was tested both neat and diluted 20-fold in single wells unless noted otherwise. After the addition of 50 μL of calibrators and samples, the plate was incubated at room temperature with shaking at 500 revolutions per minute (rpm) for 2 h. Plates were then washed with a proprietary wash buffer using an automated plate washer. A 50 μL aliquot of detection mix, including biotinylated detector antibody and buffer, was added to each well, and the plate was incubated with shaking for another hour and then washed again. For detection, a 50 μL aliquot of horseradish peroxidase–streptavidin solution was added to each well and then incubated with shaking for 30 min. After a final wash, a 50 μL aliquot of chemiluminescent substrate solution was added to each well, and the chemiluminescent intensity from the array spots in each well was immediately measured using the Q-View Imager Pro (Quansys Biosciences, Logan, UT, USA) at an exposure time of 300 s. The antigen concentrations were calculated using a five-parameter logistic fit model, which was built into the Q-View Software (Quansys Biosciences).
Luminex platform procedure
HRP2, PvLDH, and PfLDH were run in a 3-Plex Luminex assay format, and PanLDH was run separately in singleplex to avoid cross-reactivity with the species-specific pLDH assays, as described previously [13]. Briefly, following the same scheme of “Q-Plex platform procedure,” all blinded whole blood and DBS samples were assayed on the xMAP platform. Whole blood samples were directly used without any manipulation, while DBS samples were eluted by incubating a 3 mm disc punched out from each DBS in Luminex buffer ([LB] 1% bovine serum albumin in phosphate-buffered saline [PBS] with sodium azide at 0.05%), at 50 µL per punch, overnight at 4 °C in gentle agitation. Previously, sets of MagPlex® Microspheres (Luminex Corp., Austin, TX, USA) with set-specific spectral signatures were coupled with the respective monoclonal capture antibodies: anti-PvLDH (PA8, Access Bio, Somerset, New Jersey, USA), anti-PfLDH (PA11, Access Bio), anti-PANpLDH (PA12, Access Bio), anti-HRP2 (MBS832975, MyBioSource), according to the manufacture’s protocol at 30 µg/mL of microspheres.
The assay was performed by incubating 50 μL of microsphere suspension in LB (1000 microspheres of each/well in the 3-Plex or 2000 microspheres/well in the singleplex) with 50 μL of each sample, which was tested both neat (twofold final dilution per well) and diluted 20-fold in LB (40-fold final dilution per well) unless noted otherwise, overnight at 4 ºC with shaking at 600 rpm in the dark. Plates were then washed with wash buffer (0.05% Tween 20 [Sigma-Aldrich, St. Louis, MO, USA] in PBS) after first pelleting microspheres using a magnetic separator (EMD Millipore, Burlington, MA, USA). A 100 μL aliquot of detection mix containing biotinylated antibodies (EZ-Link Sulfo-NHS-Biotin Kit, Thermo Fisher Scientific, Waltham, MA, USA) to anti-PANpLDH (PA12, Access Bio) and anti-HRP2 (MBS834434, MyBioSource) for the 3-Plex assay and anti-PANpLDH (PA2, Access Bio) for the singleplex assay, all of them biotinylated in the lab using the EZ-Link Sulfo-NHS-Biotin Kit (21,435, Thermo Fisher Scientific, Waltham, MA, USA) according to manufacturer’s instructions, were added to each well at 1:1000 dilution in LB, and the plate was incubated with shaking for 1 h in the dark and then washed again. For detection, 100 µL of Streptavidin-R-Phycoerythrin (42,250, Sigma-Aldrich) at 1:1000 dilution in LB was added to each well and then incubated with shaking for 30 min in the dark. Finally, the beads were washed and resuspended in LB, and the plate was read using the Luminex xMAP 100/200 analyzer (Luminex Corp.). Fifty microspheres for each antigen were read, and the result was given as median fluorescence intensity (MFI). The antigen concentrations were calculated using a five-parameter logistic fit model using GraphPad Prism (version 6, GraphPad Software, San Diego, CA, USA). The characteristics of the xMAP and Q-Plex assay are compared in Additional file 1: Table S1.
Statistics
Whole blood and DBS eluate samples were analysed in singlet in both neat and 20-fold dilution by each assay. The working assay range was established as the range at which precision stayed under 20% of the coefficient of variation (CV) between adjusted measured concentration values of two adjacent levels. For the Q-Plex platform, the lower limit of detection (LLOD), upper limit of quantification (ULOQ), and lower limit of quantification (LLOQ) were previously described [9]. For the xMAP platform, LLOD and LLOQ were calculated using the formulas mean (blanks) + 3 × standard deviation (blanks) and mean (blanks) + 6 × standard deviation (blanks), respectively [13]. The assay characteristics including cutoff, LLOD, LLOQ, and ULOQ for HRP2, PfLDH, PvLDH, and PanLDH for the Q-Plex and xMAP platform are summarized in Additional file 1: Table S2. Since CRP assay is only incorporated into the Q-Plex platform, CRP data was not included for analysis. The dilutional linearity and inter-assay variability were determined using the WHO international standards for P. falciparum or P. vivax antigens. The dilutional linearity was assessed for the ability to generate results that have a linear response, proportional to the concentration of the international standard. The WHO international standards for P. falciparum and P. vivax antigens, ranging from 1.6 to 400 IU/mL, were used to evaluate the performance of the HRP2, PfLDH, PanLDH, and PvLDH assays with each assay-specific calibrator set. The linear range of the curve was estimated using the least squares method analyzing the regression coefficient (R2).
The inter-assay variability, a measure of the degree of reproducibility, was calculated using the CV value with a formula: (standard deviation/mean) × 100. This was determined by repeated analysis of pixel or fluorescence intensity values from identical samples of the WHO international standard P. falciparum and P. vivax antigens at 50 IU/mL, 25 IU/mL, 12.5 IU/mL, and 6.3 IU/mL concentration over multiple days for the Q-Plex and xMAP platforms, respectively. Samples were run three times over a week. The acceptable level of inter-assay variability was defined as ≤ 15% for whole blood and DBS. If necessary, the inter-assay variability was assessed using the extrapolated antigen concentration values under the same acceptance criteria (≤ 15%).
Diagnostic concordance in the identification of P. falciparum with hrp2/hrp3 deletions and Plasmodium species was estimated using the established cutoffs (mass concentrations in pg/mL). For the Q-Plex, platform cutoffs yielding 99.5% or more diagnostic specificity were used [9]. For the xMAP platform, cutoffs were calculated using the mean of negative controls + 3 standard deviations. Positive agreement (percentage) was calculated by examining the proportion of reference positive results in which the test result is positive.
Pearson correlation coefficient was used to evaluate the relationship of antigen concentrations obtained by the two platforms. Samples with concentration of antigen < lower limit of detection were excluded from the analyses. Bland–Altman plots were used to compare the difference in agreement with antigen amounts (y-axis) with the average of antigen amounts from two assays (x-axis). To assess differences in malaria antigen levels measured by the two multiplex assays, p-values were calculated using the unpaired t-test. Statistical analyses were performed using GraphPad Prism software version 6 (GraphPad Software). Any p-value of ≤ 0.05 was considered statistically significant.
Results
Linearity and reproducibility
A two-fold dilution series of the respective international standard antigen in whole blood and DBS was tested to assess linearity across the range of 0.16 to 400 IU/mL for each assay and examine the relationship between assay results and the WHO international standard antigens. All four assays (HRP2, PanLDH, PfLDH and PvLDH were linear in the range of 1.6 to 400 IU/mL on both platforms (Fig. 1). The equations for the best-fit lines are summarized for all antigens in Additional file 1: Table S3. The best-fit slope values for the respective antigen assays show a wide difference between the two platforms because the antigen quantification can be largely influenced by values assigned to the standards, but the R2 values of the curves for each of the assays were in the range of 0.9461–0.9885 and 0.9215–0.9938 for whole blood and DBS-based Q-Plex data, respectively, and 0.9037–0.9980 and 0.9796–0.9947 for whole blood and DBS-based xMAP data (Additional file 1: Table S3). Both datasets from the Q-Plex and xMAP assays demonstrated a high degree of linearity as well as correlation for all antigens.
To determine the reproducibility of the data, which should not be influenced by day-to-day variation, inter-assay performance of both platforms was measured by quantifying the CV of the assay results using the international standard antigen samples prepared in whole blood and DBS, which were tested on three different days over a week (Table 2). Overall, the minor variability in the results of the whole blood samples indicate that these two platforms were highly reproducible, with the average CV less than 10% across all platforms and all antigens. The inter-assay variability for HRP2 and PfLDH measurement using DBS samples was within acceptance range, with averages of 3.4% and 8.3% by the Q-Plex and 3.5% and 4.1% by the xMAP. The variability of the PanLDH and PvLDH results from DBS samples that were tested by the Q-Plex (Av. 26.3% and Av. 24.6%, respectively) was outside the acceptance range, whereas that from DBS samples evaluated by the xMAP were within the acceptance range (Av. 6.3% and 8.8%, respectively). However, the result of the CV calculated with PanLDH and PvLDH concentrations from the three runs showed an acceptable inter-assay variability for PanLDH (8.3%−14.5%, Av. 9.1%) and a slightly higher variability for PvLDH (5.8%–36.1%, Av. 19.0%). These results suggest overall high reproducible procedures of two platforms over time.
Identification of P. falciparum mutants with hrp2/hrp3 deletions and differentiation of Plasmodium species
To determine the effectiveness of the assay system for detecting P. falciparum mutants which fail to produce HRP2/HRP3, samples prepared with laboratory P. falciparum strains W2 (hrp2+hrp3+), Dd2 (hrp2–hrp3+), and 3BD5 (hrp2–hrp3–) were tested for the presence or absence of HRP2 and PfLDH using the established cutoff values. Both the Q-Plex and the xMAP identified P. falciparum parasites with hrp2 or hrp2/hrp3 deletions in whole blood with 100% positive agreement through evaluating the reactivity patterns for both HRP2 and HRP3, as well as PfLDH (Table 3). However, the Q-Plex appeared to be less effective than the xMAP for identification of parasites with an hrp2 deletion (80% for Q-Plex and 100% for the xMAP in detecting Dd2) and with hrp2/hrp3 deletions (40% for Q-Plex and 80% for the xMAP in detecting 3BD5) in DBS.
Both the Q-Plex and xMAP platforms detect P. falciparum in whole blood with high positive agreement (100%), with the xMAP having a higher sensitivity than the Q-Plex for P. vivax infection (Table 4). The Q-Plex shows relatively lower percent positive agreement compared to the xMAP on DBS for P. falciparum and P. vivax (Table 4). Interestingly, both the Q-Plex and xMAP misidentified P. malariae and P. knowlesi as P. falciparum and P. vivax due to cross-reactivity of pLDH from these species against PfLDH- and PvLDH-specific assays, respectively (Table 4). Notably, pLDH from these species showed different reactivity against these antibodies depending on the platforms, as demonstrated by the ratio of concentration between PanLDH to the species-specific pLDH (PanLDH/PfLDH or PanLDH/PvLDH) (Fig. 2).
Agreement between two multiplex platforms
The Pearson correlation coefficient to determine the agreement between the Q-Plex and xMAP for HRP2 showed moderate and poor positive correlations between HRP2 antigen levels in whole blood (r = 0.7432; Fig. 3) and DBS (r = 0.6432; Fig. 4), respectively. There was a notable discrepancy of the HRP2 results between the Q-Plex and xMAP assays with the dilution series of the hrp2–hrp3+ laboratory strain Dd2. In these samples, the concentration of HRP3, which may cross-react with HRP2 detector antibodies, gave a higher signal with HRP2-specific assay of the xMAP compared to the Q-Plex, which showed minimal cross-reactivity with HRP3. This is also shown in assay-specific HRP2 concentration against parasite density plots (Additional file 1: Figure S1). A systematic trend of higher assay values for PfLDH and PvLDH was observed with the Q-Plex testing in both whole blood and DBS. However, a strong agreement between the Q-Plex and xMAP was observed with PfLDH antigen, as demonstrated by Pearson correlation coefficient r values that were 0.9926 and 0.9792 for PfLDH and PvLDH in whole blood and 0.9792 and 0.9696 in DBS. As for PanLDH assay, the Pearson r value was the lowest at 0.6139 in whole blood, having two obvious clusters in a plot, but it improved to 0.8957 in DBS. The Bland–Altman plots revealed a systematic trend of difference between the Q-Plex and xMAP assays for PanLDH that was derived from the specific subsets of samples, such as P. knowlesi culture and clinical P. vivax and P. malariae, suggesting that two platforms may have different binding reactivities against pLDH proteins from these parasites (Fig. 3).
Normalization of the quantitative results against the WHO international standards
In this study, the concentration of antigen measured by two immunoassay methods was plotted against the WHO international standard antigens (Fig. 1), and the slope of each linear regression line was obtained for the “conversion factors” (Additional file 1: Table S3). Subsequently, these conversion factors were used to re-analyse the original concentration data, reporting the amount of Plasmodium antigens in IU/mL. When the normalized antigen levels in the IU/mL were compared between the Q-Plex and xMAP, there were similar distribution profiles between the two for HRP2, PfLDH, and PvLDH in both whole blood and DBS (Fig. 5). However, a statistically significant difference between the two was found for PanLDH in whole blood (p < 0.05) but not for that in DBS (p = 0.828). Similarly, correlation analysis using normalized antigen concentration values showed a strong correlation between the Q-Plex and xMAP for PfLDH and PvLDH antigens but poor to moderate correlation between the two for HRP2 or PanLDH in whole blood and DBS, demonstrating some variation in binding reactivities against HRP2/HRP3 or PanLDH (Additional file 1: Figure S2).
Discussion
Several malaria multiplex platforms have been developed to target key biomarkers, including HRP2, PfLDH, PvLDH, PanLDH, and pAldo [8, 9, 13, 22, 23]. These have proven valuable in clinical studies for understanding the biology of parasites, evaluating the performance of RDTs, and documenting the decay rates of Plasmodium antigens, among other uses. The ability to compare data across the different assays would greatly enhance the value of the results generated with these multiplexed assays. The present study sought to assess how two of these multiplex immunoassays compare in terms of the quantitative results of those Plasmodium antigens, their agreement in identifying Plasmodium species, and whether the quantitative results can be compared across platforms.
In this study, the assay performance of two malaria multiplex platforms, the Q-Plex and xMAP, which allow the quantification of four common malaria antigens—HRP2, PfLDH, PvLDH, and PanLDH—in biological samples, were compared. The xMAP assay is an in-house research assay, while the Q-Plex assay is a commercial off-the-shelf product. Both platforms are considered moderate throughput screening assays, with similar procedures and assay time to results. The xMAP provides a flexible and affordable platform by which to measure parasite antigen concentrations from a per specimen perspective once the instrumentation has been purchased, but it requires laboratories to individually source their assays reagents and validate them. The Q-Plex platform provides a commercial kit for which all components are already included and validated, but the per specimen cost is higher and there is no ability to rapidly add new assays to the platform. There are additional differences in physical characteristics, including assay format (plate-based versus bead-based), assay volume (50 µL versus 100 µL), required sample volume (12.5 µL versus 50 µL for the single target assay), and detection and reporting methods (chemiluminescence versus fluorescence, pixel intensity versus net median fluorescent intensity), which are summarized in Additional file 1: Table S1.
To assess the compatibility of data generated across both platforms, paired sample panels in whole blood and DBS were created to conduct an objective analysis of the performance of both platforms. The identical sample panels were independently analysed by two laboratories using the Q-Plex and xMAP assays. The inclusion of the WHO international standards for P. falciparum and P. vivax antigens allows standardization of the outputs of the two platforms in international units. Additionally, inclusion of hrp2/hrp3 deleted parasite strains and different Plasmodium species screened in these assay platforms allowed for a better understanding of differences in overall performance as well as the quantitative measurements derived from the two platforms. Given the differences in the assay dynamic ranges for the respective antigens between the two platforms, the sample testing was performed as blind experiments at two dilutions—neat and a second dilution. Excellent linearity was observed when using the WHO international standards in a range of 1.6–400 IU/mL and assay-specific calibrators for both the Q-Plex and xMAP assays, with R2 values > 0.9 for all antigen-specific assays. However, as demonstrated by different slope values in the linear regression fits for each analyte, the two platforms do behave differently. This is most likely because different standard materials are used to calibrate the antigen concentrations in the two platforms. Overall, the inter-assay precision expressed as CV (calculated using the signal value; acceptable ≤ 15%) were acceptable across antigens and platforms. For the Q-Plex in DBS, the PanLDH and PvLDH CV values exceeded the acceptable range and only the PanLDH signal when analysed in concentration units fell within the acceptable range. These data are indicative of a good level of reproducibility of these assays and met the predetermined assay acceptance criteria, with the Q-Plex not performing as well on DBS.
Both assays effectively identified laboratory P. falciparum parasite (3BD5) with deletion of hrp2 and hrp3 genes in addition to classifying Plasmodium species in clinical specimens using three detection reagents targeting HRP2, PfLDH, and PvLDH. In particular, the xMAP showed a higher percent positive agreement for identification of different Plasmodium infections, when the established cutoff value for each of these biomarkers was used.
PfLDH and PvLDH concentrations measured by the Q-Plex showed a strong positive correlation with those measured by the xMAP. However, HRP2 or PanLDH concentration data from the two showed a moderate to poor correlation. Overall, the Q-Plex showed 2.4-fold higher HRP2 concentration estimation compared to the xMAP. This cannot be attributed to HRP3, since a significant difference in the HRP2 assay between the two platforms is the relative response to the P. falciparum strain Dd2, which still expresses HRP3 but has an hrp2 deletion. Specifically, the Q-Plex HRP2 assay has minimal cross-reactivity with HRP3, but the xMAP assay does react strongly with HRP3. These differences are relevant to take into account when these assays are used to support RDT evaluations with clinical samples with hrp2/hrp3 deletions. Different reactivities to HRP3 have also been observed in commercially available RDTs [24, 25].
Antibodies for pLDH have very different specificity profiles such that RDTs and ELISA assays developed for PfLDH and PvLDH behave very differently across the other species of Plasmodium [26,27,28,29,30]. On the Q-Plex platform, the pLDH proteins from P. malariae and P. knowlesi were found to cross-react to PfLDH and PvLDH assays, respectively, whereas on the xMAP assay only P. knowlesi was shown to cross-react with the PvLDH assay. These differences in cross-reactivity also extend to the PanLDH signal. Substantial differences were observed between the Q-Plex and the xMAP, in the PanLDH signal as well as species-specific pLDH signal as shown by the ratio of both measurements. The PanLDH signal in the Q-Plex is calibrated against the PfLDH standard, and thus the PanLDH/PfLDH ratio is inherently 1. This study found the ratio of PanLDH/PfLDH to be close to 1 across all P. falciparum samples including the WHO international standard for P. falciparum antigen (0.805 ± 0.158), P. falciparum strains (0.892 ± 0.052–1.02 ± 0.134), and clinical samples (0.828 ± 0.119–1.068 ± 0.212). In contrast, the Q-Plex shows much higher PanLDH/PfLDH ratios with two P. malariae–infected samples (4.965 ± 0.262 and 5.61 ± 0.416), which allows the Q-Plex to differentiate these from P. falciparum infections. The PanLDH/PvLDH ratios in P. knowlesi samples (2.398 ± 0.419) tested by using the Q-Plex were similar to those from two P. vivax subsets of samples (2.318 ± 0.382, 2.26 ± 0.49). The binding patterns appear to be similar to those previously reported with the Q-Plex [31, 32]. Thus in P. vivax and P. knowlesi co-endemic populations, the Q-Plex would not be able to differentiate between the two infections without further development. Similar consistent PanLDH/PfLDH and PanLDH/PvLDH ratios for P. knowlesi and P. malariae were not observed with the xMAP, and the cross-reactivity requires further investigation. However, the xMAP can differentiate P. knowlesi from P. vivax infections due to the reduced cross-reactivity in the P. vivax assay for P. knowlesi.
The PanLDH signal also showed significant variation between the two platforms with the non–P. falciparum samples in whole blood behaving significantly differently. Overall, there was a threefold difference in PanLDH concentration as determined by the Q-Plex compared to the xMAP. These variations in PanLDH quantification most likely arise from differences in the choice of detection antibodies resulting in recognition of possibly different target epitopes and calibration standards used for the assays on each platform. These differences will impact the relative PanLDH to species-specific LDH assay signal as well as the absolute quantification of PanLDH across both assays.
The study also explored normalization of the assay signals with the WHO international standards for P. falciparum and P. vivax antigen and expression of the concentration in IU/mL. The antigen distribution plots between the two assays in quantification of HRP2, PfLDH, and PvLDH show that normalized data can be comparable between assays when the results are expressed in IU/mL. Further work is required to understand the feasibility of using international standards to normalize the PanLDH assay.
Conclusion
The data indicate that both the Q-Plex and xMAP showed good performance in detecting wild type P. falciparum, hrp2/hrp3-deleted P. falciparum mutants, and other Plasmodium species, but key differences were also observed. The study also shows agreement in quantification of HRP2, PfLDH, and PvLDH data obtained by the Q-Plex and xMAP platforms once normalization into international units with the WHO international standards for P. falciparum and P. vivax antigens is conducted. This significant finding enables comparison and utilization of results across malaria antigen quantification platforms.
Availability of data and materials
Available from the corresponding author on request.
Abbreviations
- Av.:
-
Average
- CRP:
-
C-reactive protein
- CV:
-
Coefficient of variation
- DBS:
-
Dried blood spot
- ELISA:
-
Enzyme-linked immunosorbent assay
- HRP2:
-
Histidine-rich protein 2
- HRP3:
-
Histidine-rich protein 3
- LB:
-
Luminex buffer
- LDH:
-
Lactate dehydrogenase
- LLOD:
-
Lower limit of detection
- LLOQ:
-
Lower limit of quantification
- MFI:
-
Median fluorescence intensity
- NIBSC:
-
National institute for biological standards and control
- pAldo:
-
Plasmodium aldolase
- PanLDH:
-
Pan-specific pLDH
- PBS:
-
Phosphate-buffered saline
- PfLDH:
-
P. falciparum–specific pLDH
- pLDH:
-
Plasmodium lactate dehydrogenase
- PvLDH:
-
P. vivax–specific pLDH
- R2 :
-
R squared, coefficient of determination
- RDT:
-
Rapid diagnostic test
- rpm:
-
Revolutions per minute
- SD:
-
Standard deviation
- ULOQ:
-
Upper limit of quantification
- WB:
-
Whole blood
- WHO:
-
World Health Organization
References
Mathison BA, Pritt BS. Update on malaria diagnostics and test utilization. J Clin Microbiol. 2017;55:2009–17.
Mouatcho JC, Goldring JPD. Malaria rapid diagnostic tests: challenges and prospects. J Med Microbiol. 2013;62:1491–505.
Desakorn V, Silamut K, Angus B, Sahassananda D, Chotivanich K, Suntharasamai P, et al. Semi-quantitative measurement of Plasmodium falciparum antigen PfHRP2 in blood and plasma. Trans R Soc Trop Med Hyg. 1997;91:479–83.
Gibson LE, Markwalter CF, Kimmel DW, Mudenda L, Mbambara S, Thuma PE, et al. Plasmodium falciparum HRP2 ELISA for analysis of dried blood spot samples in rural Zambia. Malar J. 2017;16:350.
Chiodini PL, Bowers K, Jorgensen P, Barnwell JW, Grady KK, Luchavez J, et al. The heat stability of Plasmodium lactate dehydrogenase-based and histidine-rich protein 2-based malaria rapid diagnostic tests. Trans R Soc Trop Med Hyg. 2007;101:331–7.
Poti KE, Sullivan DJ, Dondorp AM, Woodrow CJ. HRP2: transforming malaria diagnosis, but with caveats. Trends Parasitol. 2020;36:112–26.
Das S, Jang IK, Barney B, Peck R, Rek JC, Arinaitwe E, et al. Performance of a high-sensitivity rapid diagnostic test for Plasmodium falciparum malaria in asymptomatic individuals from Uganda and Myanmar and naive human challenge infections. Am J Trop Med Hyg. 2017;97:1540–50.
Landier J, Haohankhunnatham W, Das S, Konghahong K, Christensen P, Raksuansak J, Phattharakokoedbun P, et al. Operational performance of a Plasmodium falciparum ultrasensitive rapid diagnostic test for detection of asymptomatic infections in Eastern Myanmar. J Clin Microbiol. 2018;56:e00565-18.
Jang IK, Tyler A, Lyman C, Rek JC, Arinaitwe E, Adrama H, Murphy M, et al. Multiplex human malaria array: quantifying antigens for malaria rapid diagnostics. Am J Trop Med Hyg. 2020;102:1366–9.
Jang IK, Aranda S, Barney R, Rashid A, Helwany M, Rek JC, et al. Assessment of Plasmodium antigens and CRP in dried blood spots with multiplex malaria array. J Parasit Dis. 2021;45:479–89.
Plucinski MM, Herman C, Jones S, Dimbu R, Fortes F, Ljolje D, et al. Screening for Pfhrp2/3-deleted Plasmodium falciparum, Non-falciparum, and low-density malaria infections by a multiplex antigen assay. J Infect Dis. 2019;219:437–47.
Plucinski M, Aidoo M, Rogier E. Laboratory detection of malaria antigens: a strong tool for malaria research, diagnosis, and epidemiology. Clin Microbiol Rev. 2021;34:e0025020.
Martianez-Vendrell X, Jimenez A, Vasquez A, Campillo A, Incardona S, Gonzalez R, et al. Quantification of malaria antigens PfHRP2 and pLDH by quantitative suspension array technology in whole blood, dried blood spot and plasma. Malar J. 2020;19:12.
Plucinski MM, McElroy PD, Dimbu PR, Fortes F, Nace D, Halsey ES, et al. Clearance dynamics of lactate dehydrogenase and aldolase following antimalarial treatment for Plasmodium falciparum infection. Parasit Vectors. 2019;12:293.
van den Hoogen LL, Herman C, Presume J, Romilus I, Mondelus G, Elisme T, et al. Rapid screening for non-falciparum malaria in elimination settings using multiplex antigen and antibody detection Post hoc identification of Plasmodium malariae in an infant in Haiti. Am J Trop Med Hyg. 2021;104:2139–45.
Reichert EN, Hume JCC, Sagara I, Healy SA, Assadou MH, Guindo MA, et al. Ultra-sensitive RDT performance and antigen dynamics in a high-transmission Plasmodium falciparum setting in Mali. Malar J. 2020;19:323.
Harris LM, Campillo A, Rigsby P, Atkinson E, Poonawala R, Aidoo M, et al. Collaborative study to evaluate the proposed First World Health Organization International Standard for Plasmodium falciparum antigens. Geneva: World Health Orgnization; 2017.
Olomu C, Harris LM, Rigsby P, Atkinson E, Yerlikaya S, Ding X, et al. Collaborative study to evaluate the proposed First World Health Organization International Standard for Plasmodium vivax antigens. Geneva: World Health Orgnization; 2020.
Trager W, Jensen JB. Human malaria parasites in continuous culture. Science. 1976;193:673–5.
Moon RW, Hall J, Rangkuti F, Ho YS, Almond N, Mitchell GH, et al. Adaptation of the genetically tractable malaria pathogen Plasmodium knowlesi to continuous culture in human erythrocytes. Proc Natl Acad Sci USA. 2013;110:531–6.
Lambros C, Vanderberg JP. Synchronization of Plasmodium falciparum erythrocytic stages in culture. J Parasitol. 1979;65:418–20.
Jang IK, Tyler A, Lyman C, Kahn M, Kalnoky M, Rek JC, et al. Simultaneous quantification of Plasmodium antigens and host factor C-reactive protein in asymptomatic individuals with confirmed malaria by use of a novel multiplex immunoassay. J Clin Microbiol. 2019;57:e00948-e1018.
Woodford J, Collins KA, Odedra A, Wang C, Jang IK, Domingo GJ, et al. An experimental human blood-stage model for studying Plasmodium malariae infection. J Infect Dis. 2020;221:948–55.
Kong A, Wilson SA, Ah Y, Nace D, Rogier E, Aidoo M. HRP2 and HRP3 cross-reactivity and implications for HRP2-based RDT use in regions with Plasmodium falciparum hrp2 gene deletions. Malar J. 2021;20:207.
Das S, Peck RB, Barney R, Jang IK, Kahn M, Zhu M, et al. Performance of an ultra-sensitive Plasmodium falciparum HRP2-based rapid diagnostic test with recombinant HRP2, culture parasites, and archived whole blood samples. Malar J. 2018;17:118.
Hurdayal R, Achilonu I, Choveaux D, Coetzer TH, Goldring JPD. Anti-peptide antibodies differentiate between plasmodial lactate dehydrogenases. Peptides. 2010;31:525–32.
Tomar D, Biswas S, Tripathi V, Rao DN. Development of diagnostic reagents: raising antibodies against synthetic peptides of PfHRP-2 and LDH using microsphere delivery. Immunobiology. 2006;211:797–805.
Foster D, Cox-Singh J, Mohamad DS, Krishna S, Chin PP, Singh B. Evaluation of three rapid diagnostic tests for the detection of human infections with Plasmodium knowlesi. Malar J. 2014;13:60.
McCutchan TF, Piper RC, Makler MT. Use of malaria rapid diagnostic test to identify Plasmodium knowlesi infection. Emerg Infect Dis. 2008;14:1750–2.
Grigg MJ, William T, Barber BE, Parameswaran U, Bird E, Piera K, et al. Combining parasite lactate dehydrogenase-based and histidine-rich protein 2-based rapid tests to improve specificity for diagnosis of malaria due to Plasmodium knowlesi and other Plasmodium species in Sabah. Malaysia J Clin Microbiol. 2014;52:2053–60.
Barney B, Velasco M, Cooper C, Rashid A, Kyle D, Moon R, et al. Diagnostic characteristics of lactate dehydrogenase on a multiplex assay for malaria detection including the zoonotic parasite Plasmodium knowlesi. Am J Trop Med Hyg. 2021;106:275–82.
Kho S, Anstey NM, Barber BE, Piera K, William T, Kenangalem E, et al. Diagnostic performance of a 5-plex malaria immunoassay in regions co-endemic for Plasmodium falciparum, P vivax, P knowlesi, P malariae and P ovale. Sci Rep. 2022;12:7286.
Acknowledgements
The authors thank David Boyle for providing feedback on the manuscript and Sarah Gutema for the manuscript proofing and editing.
Funding
This work was supported by the Bill & Melinda Gates Foundation (grant number INV1135840). AM is supported by the Departament d’Universitats i Recerca de la Generalitat de Catalunya, Agència de Gestió d’Ajuts Universitaris i de Recerca (2017SGR664). The Centro de Investigaçao em Saude de Manhica (CISM) is supported by the Government of Mozambique and the Spanish Agency for International Development Cooperation (AECID). We also acknowledge support from the Spanish Ministry of Science and Innovation through the Centro de Excelencia Severo Ochoa 2019–2023 Program (CEX2018-000806-S) and support from the Generalitat de Catalunya through the CERCA Program. This research is part of ISGlobal’s Program on the Molecular Mechanisms of Malaria, which is partially supported by the Fundación Ramón Areces.
Author information
Authors and Affiliations
Contributions
Conceived: AG, AM, GJD, IKJ. Performed the experiments: AJ, AR, RB. Managed the data: XCD. Analysed the data: AJ, IKJ. Wrote the paper: AM, GJD, IKJ. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
All authors declare no competing financial interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Additional file 1:
Table S1. Comparison of two multiplex platforms for the quantitative assessment of malaria antigens. Table S2. Characteristics of the Q-Plex and xMAP assays. Table S3. Linear relation between the WHO international standard P. falciparum and P. vivax antigens and assay concentration results in whole blood or DBS. Regression analysis showed the fit of the data with R2 values that met the acceptance criteria (R2 ≥ 0.85). One operator conducted three experiments over a week. Figure S1. HRP2 concentration in (A) whole blood and (B) DBS plotted against the parasitemia. Three subsets of dilution-series samples composed of different P. falciparum laboratory strains (W2, Dd2, and 3BD5) were tested by using the Q-Plex and xMAP. Measured concentration was plotted on the y-axis against parasitemia (par/mL) on the x-axis. Red and blue symbols and regression lines represent samples spiked with different strains and test types. Figure S2. Correlation analysis. Correlation plots of antigen concentration corrected against the WHO international standard P. falciparum and P. vivax antigens, P. falciparum culture, and clinical samples in (A) whole blood and (B) DBS. The Pearson correlation r value and the equation of the regression line obtained from analysis using log transformed data are shown. Outlier sample sets were color-coded.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Jang, I.K., Jiménez, A., Rashid, A. et al. Comparison of two malaria multiplex immunoassays that enable quantification of malaria antigens. Malar J 21, 176 (2022). https://doi.org/10.1186/s12936-022-04203-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12936-022-04203-9